摘要
使用自行研制的全自动流动注射分析仪,采用在线蒸馏流动注射4-氨基安替比林显色法测定水中的挥发酚。方法的检出限为0.58μg/L;线性范围为2~50μg/L,线性相关系数r为0.9994;其相对标准偏差为4.1%~6.2%;实际水样的加标回收率为94%~104%;样品测定频率为9样/小时。本方法快速、简便、灵敏度较高,线性范围、精密度和准确度均符合国家标准。
Volatile phenolics in water were determined by using flow injection analysis - 4 - AAP method using auto- mated flow injection analyzer developed by ourself. The detection limit was 0. 58μg/L and the liner range was 2 - 50μg/L (r = 0. 9994). The relative standard deviation was 4. 1% -6. 2% and the recovery was 94% - 104%. The method is rapid, simple, sensitive, the linear range, precision and accuracy all consist with GB standard.
出处
《现代仪器》
2007年第3期69-71,73,共4页
Modern Instruments
基金
国家科技部"十五"科技攻关项目(编号:2004BA210A05)
关键词
流动注射分析
在线蒸馏
挥发酚
Flow injection analysis On- line distillation Volatile phenolics