期刊文献+

微创介入手术主动导管的矩阵式驱动系统的设计

Study on an Actuation System for Matrix Control of the Active Catheter in a Minimally-lnvasive Intervention Surgery
在线阅读 下载PDF
导出
摘要 根据徽创介入手术主动导管尺寸微小、空间有限不能容纳过多导线的结构特点,提出了一种矩阵式驱动控制系统,使用最少的导线完成对多个SMA驱动器的控制。利用单片机输出的脉宽调制(PWM)信号对脉冲电流进行调节和多个驱动器之间的相互配合作用,实现主动导管的导向动作。 As it is impossible for an active catheter with a very small space to accommodate overmany lead wires in minimally-invasive surgery, a matrix network system is presented, in this paper, to control SMA actuators using minimum lead wires. Pulse current is adjusted by pulse width modulation (PWM) signals from the single-chip processor. In addition, multiple SMA actuators' cooperation helps the active catheter to succeed in guiding motion.
出处 《中国医疗器械杂志》 CAS 2006年第6期416-418,430,共4页 Chinese Journal of Medical Instrumentation
基金 黑龙江省科技攻关重点资助项目(GC04A510)。
关键词 主动导管 形状记忆合金 矩阵式控制系统 脉宽调制 active catheter, shape memory alloy, matrix actuation system, pulse width modulation (PWM)
  • 相关文献

参考文献8

  • 1G.T.Gillies,FL C.Ritter,W.C.Broaddus,et al.Magnetic manipulation instrumentation for medical physics research.Rev.Sci.Inst.,1994,3 (65):533-562
  • 2Y.L.Fu,X.L.Li,Z.G.Liang,et al.Magnetic Manipulation System for Intravascular Catheterization.The 12th International Conference on Biomedical Engineering.2005
  • 3S.Guo Fukuda,T.Fukuda and F.ARAI.Micro active guide wire catheter system.IEEE International Conference on Intelligent Robots and Systems.1995,2:172-177
  • 4S.Guo,T.Nakamura and T.Fukuda.Micro Active Guide Wire Catheter Using ICPF Actuator.International Workshop on Advanced Motion Control.1996,2:729-734
  • 5Y.Haga,Y.Tanahashi,M.Esashi.Small diameter active catheter using shape memory alloy.IEEE MEMS workshop.1998:419-424
  • 6Yoichi Haga,Takashi Mineta,Masayoshi Esashi.Multi-functional Active Catheter.Sensor Update 2000.8,(1),11:147-186
  • 7Ki-Tae Park,Masayoshi Esashi.A Multilink Active Catheter with Polyimide -Based Integrated CMOS Interface Circuits.Journal of Micro-electromechanical System.1999,12:349-356
  • 8Hyo Jik Lee,Jung Ju Lee,Dong Soo Kwon et al.Neural network based control of SMA actuator for the active catheter.Korea Science and Engineering Foundation.2000:1-6

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部