摘要
目的配制血糖即时检验室间质控品并做评价。方法准备肝素抗凝全血标本,取部分全血即刻分离血浆在日立7600-110全自动生化仪上测定作为对照组,剩余的全血联合加入多种抗糖酵解剂,并在室温(22OC)保存2、8、24、48小时后测定血中葡萄糖的浓度,在日立7600-110全自动生化仪、雅培MediSense○ROptiumTM快速血糖仪、强生SureStepTMPlus稳步型血糖仪、罗氏GLUCOTREND○R2乐康全血糖仪上测定,并将各仪器各时间段的结果作组间比较。结果含有复合抗酵解剂的全血标本,可使血中葡萄糖浓度在24小时内保持稳定,24小时内各时间段之间的结果差异无显著性意义(F=1.52,P>0.05),48小时的血糖结果有所下降,与对照组结果比较,有显著性差异(P<0.01)。结论自制的全血血糖室间质控品配制简单,血糖浓度24小时内保持稳定,且不会干扰快速血糖仪的测定,符合血糖即时检验室间质控要求。
Objective:To confect and evaluate the external quality control sample of Glucose concentration detection in point of care testing. Methods:Collected whole blood specimens with heparin as anticoagulant , took a part of whole blood to separate plasma which would be determined by HITACHI 7600-110 Automatic Analyzer for control group,added many kinds of antiglycolytic agents in different concentrations to the remanent blood, preserved at room temperature (22℃),and determined blood glucose concentration at different time points (2,8,24,48h) by HITACHI 7600-110 Automatic Analyzer , MediSenseOptium^TM Blood Glucose Electrodes , LifeScan SureStep^TM Plus glucometer,Roche GLUCOTREND2 glucometer ,then compared with the results of glucose concentration detection at different time points. Results: the blood glucose concentration of whole blood specimens which contained mixture antiglycolytic agents kept stable ,the results at different points in 24h didn't have prominent significances(F= 1.52,P〉0.05) . The blood glucose concentrations declined after 48h , comparing with the results of control group, had prominent significances (P〈0. 01). Conclusion: confecting the self-regulating whole blood external quality control sample is simple , glucose concentration of the whole blood external quality control sample can keep stable in 24h , furthermore the sample don't interfere the determining of glucose concentration by some glucometers that use extensively in the hospital POCT , the self-regulating whole blood external quality control sample accord with the request of the external quality control sample.
关键词
即时检验
快速血糖仪
室间质控品
制备
评价
Point-of Care Testing
Glucometer
the external quality control sample
preparation
evaluation