期刊文献+

实验仪器测试平台的抗振设计

Anti-vibration design for experiment instrument testing bench
在线阅读 下载PDF
导出
摘要 实验仪器测试平台是准确测量的重要物质基础,它受到的振动可归结为单自由度的振动问题。通过对单自由度振动系统特性分析,分析相关抗振因子特性,从而得出实验测试平台的抗振设计基本原理,即只有当外激频率fj大于系统频率f0的2倍时,抗振系统才有效;进而从材料、结构及措施上进行抗振介绍,以全息照相实验平台的抗振动设计为例作了测算和设计。 Experiment instrument testing bench is important matter foundation for accurate measurement, Vibration comes from a single variance. This paper analyzed system feature and correlative anti-vibration factor of the single variance and came to the conclusion of anti-ibration design basic principle for experiment instruent testing bench, Anti-vibration system is valid when exciting frequecy is greater than √2 multiple system inherent frequency. This paper also introduced antl-vibration design for material, structure and finally, gave the calculation and design taking holograph experiment testing bench as example.
作者 蓝军
出处 《中国测试技术》 CAS 2006年第3期56-58,共3页 CHINA MEASUREMENT & TESTING TECHNOLOGY
关键词 实验仪器设备 测试平台 单自由度 振动系统 全息照相 抗振设计 Experiment instrument Testing center Single variance Vibratlonal system Holograph Anti-vibration design
  • 相关文献

参考文献6

二级参考文献11

  • 1[1]Koliopoulos C L. Simultaneous phase shift interferometer. SPIE, 1991, 1513:119~127
  • 2[2]Millerd J E, Brock N J. U. S. Patent 6, 304, 330, 2001
  • 3[3]Kwon O Y. Multichannel phase-shifted interferometer. Opt. Lett., 1984, 9(12):59~61
  • 4[4]Colucci D, Wizinowich P. Millisecond phase acquisition at video rates. Appl. Opt., 1992, 31(28): 5919~5925
  • 5[5]North-Morris M B, van Delden J,Wyant J C. Phase-shifting birefringent scatterplate interferometer. Appl.Opt,2002,41(4):668~677
  • 6[6]Freschi A A, Frejlich J. Adjustable phase control in stabilized interferometry. Opt. Lett., 1995, 20(6): 635~637
  • 7[7]Yamaguchi I, Liu J Y, Kato J. Active phase-shifting interferometers for shape and deformation measurements. Opt. Eng.,1996,35(10): 2930~2937
  • 8[8]Cole G C, Burge J H, Dettmann L R. Vibration stabilization of a phase shifting interferometer for large optics. SPIE, 1997,3134:438~446
  • 9[9]Zhao Chunyu, Burge J H. Vibration-compensated interferometer for surface metrology. Appl. Opt., 2001, 40(34): 6215~6222
  • 10孟庆丰,屈梁生.Wigner分布及其在机械故障诊断中的应用[J].信号处理,1990,6(3):155-162. 被引量:16