期刊文献+

高精度中远红外辐射定标技术研究 被引量:5

High Accurate Calibration of Optical Radiation at Middle Infrared and Far Infrared Spectrum
在线阅读 下载PDF
导出
摘要 红外辐射定标是红外遥感信息定量化的关键技术之一。从我国空间红外传感器的发展要求出发,结合国外基于绝对低温辐射计的红外辐射定标技术研究状况,提出了利用宽波段可调谐激光器和球状InSb、HgCdTe光电探测器以及薄膜热电堆探测器TS-76来实现高精度中远红外辐射定标的技术方案。 Infrared radiometric calibration is of critical importance for information quantification of satellite remote sensing of environment at infrared spectrum. In this paper we propose plans for laboratory calibration at middle infrared and far infrared spectrum for Chinese space sensors. We review and compare various advanced calibration techniques, and suggest calibration methods based on widely tuneable lasers, HgCdTe/sphere detectors, InSb/sphere detectors and thin -film thermopiles TS -76. Technical proposals are briefly discussed.
出处 《红外技术》 CSCD 北大核心 2006年第3期178-183,共6页 Infrared Technology
关键词 绝对低温辐射计 红外辐射定标 不确定度 cryogenic absolute radiometer infrared radiometric calibration uncertainty
  • 相关文献

参考文献18

  • 1Evangelos Theocharous,Frank J.J.Clarke,Leon J.Rogers,Nigel P.Fox.The latest measurement techniques at NPL for the characterisation of infrared detectors and materials[A].SPIE[C].2003,5209:228~239.
  • 2Ohno Y.Detector-based luminous-flux calibration using the Absolute Integrating-Sphere Method[J].Metrologia.1998,35(4):473~478.
  • 3Theocharous E,Prior T R,Haycocks P R,Fox N P.High-accuracy,infrared,spectral responsivity scale[J].Metrologia.1998,35(4):543~548.
  • 4Schrama C A,Bloembergen P,E W M van der Ham.Monochromator -based cryogenic radiometry between 1μm and 20 μm[J].Metrologia.2000,37(5):567~570.
  • 5Bridges J M,W R Ott.Vacuum ultraviolet radiometry.3:The argon mini-arc as a new secondary standard of spectral radiance[J].Appl Opt.1977,16(2):367~376.
  • 6Bridges J M,Migdall A L.Characterization of argon arc source in the infrared[J].Metrologia.1995,32 (6):625~628.
  • 7Alan L.Migdall,George P.Eppeldauer.NIST MEASUREMENT SERVISES:Spectrorndiometric Detector Measurements:Part Ⅲ-Infrared Detectors[M].December.1998.
  • 8Migdall A,Eppeldauer G.Realization of an infrared spectral radiant power response scale on a cryogenic bolometer[J].Metrologia.1998,35(4):307~315.
  • 9Rice J P.An electrically substituted bolometer as a transfer-standard detector[J].Metrologia.2000,37 (5):433~436.
  • 10Lorentz S R,Ebner S C,Walker J H,Datla R U.NIST Low-background Infrared spectral calibration facility[J].Metrologia.1995,32(6):621~624.

二级参考文献14

  • 1T J Quinn, J E Martin. Cryogenic Radiometry and Prospects for Further Improvements in Accuracy [J]. Metrolngia, 1991, 28(3): 155-161.
  • 2C C Hoyt, P V Foukal. Cryogenic Radiometers and their Application to Metrology[J]. Metrologia, 1991, 28(3): 163-167.
  • 3P V Foukal, H Kochling, P Miller. Cryogenic absolute radiometer as laboratory irradiance standards, remote sensing detector and pyroheliometers[J]. Appl Opt, 1990, 29(7): 988-993.
  • 4N P Fox. Trap Detector and their Properties[J ]. Metrologia,1991, 28(3): 197-202.
  • 5P Karha, A Haapalinna, P Toivanen, et al. Filter radiometry based on direct utilization of trap detectors[J ]. Metrologia, 1998, 35(4): 255-259.
  • 6E G Atkinson. Butler Calibration of an InGaAs photodiode at 1300nm with a cryogenic radiometer and a diode laser[J ]. Metrologia, 1998, 35(4): 241-245.
  • 7B N Taylor, C E Kuyatt. Guidelines for evaluating and expressing the uncertainty of NIST measurement results[J/OL]. http: //physics. nist. gov/pubs/guidelines.
  • 8Kostkovski H J.Reliable Spectroscopy[].Maryland: Spectroradiometry Consulting.1997
  • 9Hoyt C C.Foukal P V Cryogenic radiometers and their application to metrology[].Metrologia.1991
  • 10Gentile T R,Houston J M,Hardis J E,et al.The NIST high accuracy cryogenic radiometer[].Applied Optics.1996