一种新型中子周围剂量当量(率)测量装置的性能实验 被引量:5

PERFORMANCE TEST OF A NEW NEUTRON MEASURING APPARATUS FOR AMBIENT DOSE EQUIVALENT(RATE)
在线阅读 下载PDF
导出
摘要 本文介绍一种新型中子周围剂量当量(率)监测方法,搭接了一套单球多计数器的中子周围剂量当量(率)监测装置,利用蒙特卡罗MCNP4A程序模拟计算了中子注量的响应函数,借用“少道”解谱软件进行解谱并对一些辐射防护量进行了计算。通过对252Cf和Am-Be两种中子参考辐射场的测试,表明对中子注量率、周围剂量当量率、单位注量平均周围剂量当量和剂量当量平均能量的计算结果与参考值相比,偏差均小于10%,这种监测方法较好地解决了中子剂量仪表的能量响应问题。 Based on a new monitoring method for neutron ambient dose equivalent (rate) introduced in this report, a neutron measuring apparatus for the ambient dose equivalent (rate) witha single sphere and multicounters was set up. The fluence energy response functions were simulated calculated by Monte Carlo MCNP4A program, some radiation protection quantities were evaluated utilizing “few-channel” unfolding code. Through the test for reference radiation fields of ^252Cf and Am-Be, it shows that deviations of calculated results of the neutron fluence rate, ambient dose equivalent rate, per fluence average ambient dose equivalent and dose equivalent average energy are less than 10% comparing to their reference values. This monitoring method could better improve the energy response of neutron dosemeters.
出处 《辐射防护》 CAS CSCD 北大核心 2005年第5期283-290,共8页 Radiation Protection
关键词 中子位置灵敏计数器 中子周围剂量当量 能量响应矩阵 少道解谱 辐射防护 Neutron Position-sensitive Counter, Neutron Ambient Dose Equivalent, Energy Response Matrix, Few-channel Unfolding
  • 相关文献

参考文献24

  • 1Ing H, Clifford T, et al. ROSPEC-A Simple Reliable High Resolution Neutron Spectrometer. Radi. Prot. Dose.,1997.70(1-4):273
  • 2Toyokawa H, Uritani A,Mori C. Neutron Spectrometer with Position-sensitive Proportional Counters. Nucl. Instrum. and Meth., 1996. A381:481
  • 3Toyokawa H, Uritani A,Mori C. A Multipurpose Spherical Counter. IEEE Transaction on Nuclear Science, 1995. 42(4):28
  • 4Liu J C. Evalution of a Mcrodosimetric-based Neutron Instrument. Radi. Prot. Dosi., 2000. 88(4):323
  • 5Uritani A, Kuniya Y,Kobayashi K. A Long and Slender Position Sensitive Helium-3 Proportional Counter with An Anode Wire Supported by A Ladder-shaped Solid Insulator. Journal of Nucl. Sci. and Tech., 1995. 32(8):719
  • 6Matoba M, Sakae T,Yamazaki T. Characteristics and Construction of Charge-division Position-sensitive Proportional Counter. Nucl. Instrum. and Meth., 1982. A196(1):257
  • 7MCNP4A. Monte Carlo N-particle Transport Code System. Los Alamos National Laboratory. Los Alamos: New Mexico,1993
  • 8李桃生,杨连珍,李冬馀.单球多计数器的中子能量响应计算[J].核电子学与探测技术,2004,24(2):132-135. 被引量:9
  • 9李桃生,杨连珍,陈军,马吉增,侯雪莉.单球多计数器中子解谱方法和剂量评价简析[J].核电子学与探测技术,2004,24(5):453-456. 被引量:8
  • 10Snieder R,Scales J A. Are Inverse Problem III-posed?Department of Geophysics and Center for Wave Phenomena Colorado School of Mine. CO80401.USA,2000

二级参考文献38

  • 1陈常茂 等.多球中子探测器能量响应特性的研究[J].辐射防护,1983,3(1):14-14.
  • 2Aroua A, et al. Evaluation and test of the response matrix of a multisphere neutron spectrometer in a wide energy range, part Ⅱ Simulation[J].Nucl Instrum Methods, 1992, A321(2) :333.
  • 3Nolan EH, et al.. The response of Bonner spheres ro neutrons from thermal energy to 17. 3MeV[J].Nucl Instrum Methods, 1985, A238(3) :509.
  • 4Aroua A et al. Improved neutron spectrometerbased on bonner spheres [J]. Radi Prot Dosi,1997, 70(1-4):285.
  • 5Liu JC,et al. Neutron spectral measurements at ORNL[J]. Radi Prot Dosi 1990, 30(3):169.
  • 6Karny M, et al. Bayes-based solution of dosimteric problem and related unfolding to bonner spectrometer[J]. Radioisotopy, 1989,30(5) :350.
  • 7MCNP4A. Monte Carlo N-particle transport code system. Los Alamos national laboratory, Los Alamos, New Xexico, 1993.
  • 8Wiegel B, et al. Calculations of the response functions of bonner spherical 3He proportional counter using a realistic detector model[J]. PTB-Bericht N-21, 1994.
  • 9Akira U,et al. Development of position helium-3 proportional counter for thermal neutrons with novel signal readout[J]. Journal of Nucl Sci Tech,1990, 27(8): 712.
  • 10Toyokawa H, et al. A spherical neutron counter with an extended energy response for dosimetry[J]. Radi Prot Dosi, 1997. 70(1-4):365.

共引文献14

同被引文献64

  • 1李桃生,杨连珍,陈军,马吉增,侯雪莉.单球多计数器中子解谱方法和剂量评价简析[J].核电子学与探测技术,2004,24(5):453-456. 被引量:8
  • 2朱传新,陈渊,牟云峰,郭海萍,王新华,安力.钨-铜合金影锥屏蔽体屏蔽性能的MCNP程序计算[J].原子能科学技术,2006,40(1):21-25. 被引量:2
  • 3李宗强,苏有武,朱小龙,李武元,王桂玲,伏明灯.一个扩展量程球形中子剂量仪的研制[J].核电子学与探测技术,2006,26(6):939-941. 被引量:1
  • 4赵士庵,欧向明.我国辐射防护仪器的应用现状和校准[J].中国辐射卫生,2006,15(4):409-411. 被引量:6
  • 5丁大钊 叶春堂 赵志祥 等.中子物理学[M].北京:原子能出版社,2001.54-59.
  • 6国际放射防护委员会.ICRP74号报告,外照射放射防护中使用的换算系数[R].1998.
  • 7ISO. Neutron reference radiation for calibrating neu- tron - measuring devices used for radiation p rotec- tion purposes and for determining their response as function of neutron energy [ P] . ISO 8529, 1989.
  • 8Recommendations of the international commission on radiological protection, ICRP Publication 60 [ R ]. Ox-.
  • 9IEC1332 ( 1994 ). International Standard. Radiation Rrotection Instmmention - Installed Dose Equivalent Rate Meters, Warning Assemblies and Monitors for Neutron Radiation Of Energy from Thermal to 15 MeV [S].
  • 10M. Mocko,L. L. Daemen, M. A. Hartl,et al.. Experi- mental study of potential neutron moderator materials [ J ]. Nuclear Instruments and Methods in Physics Re- search A,2010,624 : 173 - 179.

引证文献5

二级引证文献5

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部