对比剂注射速率对脑CT灌注成像的影响

An experimental study on contrast medium injection rate in cerebral CT perfusion imaging
暂未订购
导出
摘要 目的探讨多层螺旋CT脑灌注成像合适的注射速率。方法实验采用配伍组设计,处理因素为对比剂注射速率,分别为3、4、5、6、8ml/s;配伍因素为对比剂注射剂量,分别为10、15、20、25、30ml。25只比格犬随机入组进行灌注成像扫描。所得图像经后处理,测量上矢状窦内对比剂至峰值时间及升高CT值。结果注射速率对上矢状窦至峰值时间有统计学意义。与8ml/s组比较,5、6ml/s组无统计学意义,3、4ml/s组上矢状窦至峰值时间延长;注射速率对上矢状窦峰时升高CT值无统计学意义。结论注射速率低至5ml/s脑CT灌注成像亦可获得满意效果。 Objective To investigate the effect of suitable injection rate on multislice CT perfusion imaging. Methods Randomized block design was adopted in this study. Injection rates of contrast medium were 3, 4, 5, 6 and 8 ml/s,and the dosages of that 10, 15, 20, 25 and 30 ml. 25 Beagle dogs were randomly enrolled to undergo cerebral CT perfusion imaging. After postprocessing, the time to peak (TTP) and risen CT value (RV)of superior sagital venous were measured. Results Injection rate had a statistical significant effect on TTP of superior sagital venous. There was no significant difference among group 5, 6 and 8 ml/s. Compared with 8 ml/s, TTP of superior sagital venous was prolonged when injection rates of 3, 4 ml/s were adopted. Injection rate had no statistical significant effect on RV of superior sagital venous. Conclusion The quality of cerebral CT perfusion imaging is good enough even in an injection rate as low as 5 ml/s.
出处 《江苏医药》 CAS CSCD 北大核心 2005年第9期672-674,i0002,共4页 Jiangsu Medical Journal
基金 江苏省"135"医学重点人才工程项目(135-55)
  • 相关文献

参考文献7

  • 1Eastwood JD, Lev MH, Provenzale JM. Perfusion CT with iodinated contrast material. Am J Roentgenol, 2003,180: 3-12.
  • 2Richard E, Howard Y,George JH,et al. Guidelines and recommendations for perfusion imaging in cerebral ischemia: a scientific statement for healthcare professionals by the writing group on perfusion imaging, from the Council on Cardiovascular Radiology of the American Heart Association. Stroke, 2003,34:1084-1104.
  • 3Koenig M, Klotz E, Luka B, et al. Perfusion CT of the brain:diagnostic approach for early detection of ischemic stroke. Radiology, 1998,209: 85-93.
  • 4Eastwood JD, Provenzale JM, Hurwitz LM, et al. Practical injection-rate CT perfusion imaging: deconvolution-derived hemodynamics in a case of stroke. Neuroradiology, 2001, 43: 223-226.
  • 5Meier P, Zierler KL. On the theory of the indicator-dilution method for measurement of blood flow and volume. J Appl Physiol, 1954,6:731-744.
  • 6Hoeffner EG, Case I, Jain R, et al. Cerebral perfusion CT: technique and clinical applications. Radiology, 2004,231: 632-644.
  • 7赵虹,周康荣,严福华.肝脏多层螺旋CT扫描:不同注射流率对肝脏对比剂循环时间影响的探讨[J].中华放射学杂志,2004,38(1):82-85. 被引量:37

二级参考文献1

  • 1Joern J. Sandstede,Alexander Tschammler,Meinrad Beer,Carsten Vogelsang,Guenther Wittenberg,Dietbert Hahn. [J] 2001,European Radiology(8):1396~1400

共引文献36

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部