期刊文献+

“患者”——伴有超滤人工肾系统的动力学模型

DYNAMIC MODEL IN SYSTEM OF PATIENT-ARTIFICIAL-KIDNEY WITH ULTRAFILTRATION
暂未订购
导出
摘要 本研究从理论上建立了“患者”-中空纤维型人工肾系统传质动力学的单室模型。该模型考滤了超滤对溶质清除率及体液体积的影响,具有较为普遍的意义。向犬体注入尿素和肌酐代替患者进行血液透析的活体实验,经对实验数据进行拟合,获得了模型参数G和V_B(0),并与生理值进行了比较。 <ABSTRACT> This paper established theoretically a single compartment model of mass transfer dynamics in the system of patient-hollow-fiber artificial kidney with ultrafiltration. Because the model considered the influence of ultrafiltration on solute clearance as well as body fluid volume change, it has universal significance. In author's experiment, in vivo experiments were perfomed by injecting urea and creatine into dogs pbody. By fitting the experimental data, the model parameters G and VB (0) were obtained as well as useful conclusion after comparing the model parameters with the physiological values.
出处 《中国生物医学工程学报》 EI CAS CSCD 北大核心 1989年第4期250-257,共8页 Chinese Journal of Biomedical Engineering
基金 国家自然科学基金资助项目
关键词 人工肾 动力学模型 血液透析 Mass Transfer Ultrafiltration