摘要 目的:评价电子血压计测量血压的准确性。方法: 200名体检者用BP-203RVⅡ型电动血压计同时与台式汞柱血压计测量血压,进行自身对照并比较两种测量结果。结果:两种血压测量结果差异无显著性 (P>0.05)。结论:BP-203RVⅡ型电动血压计测定血压可靠、方便、客观,有良好的应用前景。 Objective: To appraise the accuracy of electronic sphygmomanometer in measuring the blood pressure. Method: 200 physical examinee were measured blood pressure with the desk-top mercury column sphygmomanometer at the same time with the BP-203RV Ⅱ type electronic sphygmomanometer, carried on one's own contrast and compare the two kinds of measurement results. Result: There was no significant difference between the two kinds of results. Conclusion: BP-203RV Ⅱ type electronic sphygmomanometer is reliable, convenient, objective to determine blood pressure, and there is good application prospects.
机构地区 深圳市中医院
出处 《齐鲁护理杂志》 2005年第1期21-22,共2页 Journal of Qilu Nursing
关键词 血压测定 血压计 随机对照实验 the blood pressure determining sphygmomanometer randomized contrast experiment