INFLUENCE OF MIMIC CARDIAC RATE ON HYDRODYNAMICS OF DIFFERENT MECHANICAL PROSTHETIC CARDIAC VALVES IN VITRO 被引量:2

INFLUENCE OF MIMIC CARDIAC RATE ON HYDRODYNAMICS OF DIFFERENT MECHANICAL PROSTHETIC CARDIAC VALVES IN VITRO
暂未订购
导出
摘要 Objective To assess the influence of mimic cardiac rate on hydrodynamics of different mechanical prosthetic cardiac valves. Methods US-made CarboMedics bileaflet valve, China-made Jiuling bileaflet valve and C-L tilting disc valve were tested via a pulsatile flow simulator in the aortic position. Testing conditions were set at mimic cardiac rates of 55 bpm, 75 bpm, 100 bpm with a constant mimic cardiac output of 4 L/min. The mean pressure differences (ΔP), leakage volumes (LEV) and closing volumes (CLV) across each valve, and effective orifice areas (EOA) were analyzed. Results Within physiological range, ΔP, LEV, and CLV decreased as mimic cardiac rate increased, with a large extent of variance. EOA increased along with an increase in mimic cardiac rate. It was a different response in terms of cardiac rate alteration for different types of mechanical prosthetic cardiac valves. Conclusion Mimic cardiac rate change affects hydrodynamics of mechanical prosthetic cardiac valves. Within physiological range, the hydrodynamic of prosthetic bileaflet valve is better than that of tilting disc valve.
出处 《Chinese Medical Sciences Journal》 CAS CSCD 2005年第2期138-141, ,共4页 中国医学科学杂志(英文版)
关键词 mimic cardiac rate mechanical prosthetic cardiac valves HYDRODYNAMICS 心率 流体力学 心导管 病理机制