摘要
本文用CS-9000薄层扫描仪同时测定了硝西泮、氯硝西泮、安定、艾司唑仑、安眠酮、氯氮、三唑仑等7种镇静催眠药物的血药浓度,以正庚烷:氯仿:无水乙醇:乙酸乙酯(5:2:2:1.5)为展开剂,Rf值分别为0.68.0.63,0.59,0.34,0.71,0.49,0.28,测定波长分别为260nm,240nm,31Onm,238nm,305nm,264nm,230nm,线性范围1~10μg/ml,相关系数0.9972~0.9999,回收率:92.86%~100.6%,板内板间误差CV%0.28%~4.92%。
edative-hypnotics are used extensively in clinical therapy.Sometimes people took an over dosage or committed suicide bytakinging these drugs. Doctors did not know which kind of these drugs causing poising.In order to give an early diagnosis foreffective rescue,we have established a rapid,accurate and reliable TLC methed. Seven sedative hypnotics could be determinedsimultaneously using one developing system. The clinical application was satisfactory.
出处
《中国医院药学杂志》
CAS
CSCD
北大核心
1994年第1期3-5,共3页
Chinese Journal of Hospital Pharmacy