摘要
本文从理论上分析了温度、粘度对涡轮流量计测量结果的影响,介绍了用水标定的流量计用于 LH_2、LOX 流量测量的实际结果。用水标定的涡轮流量计用于低温液体流量测量引起的附加误差主要是由于温度和粘度不同引起的。
This paper theoreticly analyses the influence of temperature and viscosity upon the measurement result of terbine flowmeter,presents the acturl experiment result which was gained by terbine flowmeter with LH2 and LOX. These flowmeters used for measurement were calibrated with water.
出处
《低温与超导》
CAS
CSCD
北大核心
1989年第2期7-10,共4页
Cryogenics and Superconductivity