期刊文献+

催化动力学光度法测定痕量铜的研究 被引量:27

Catalytic Kinetic Spectrophotometric Determination of Trace Copper(Ⅱ)
在线阅读 下载PDF
导出
摘要 本文研究了铜(Ⅱ)催化抗坏血酸还原DCB偶氮胂的褪色反应,测定了反应级数和表观活化能,探讨了反应机理,建立了测定痕量铜(Ⅱ)的新方法。本法检出限为7.6×10 ̄(-12)g/mL,线性范围为0~60ng/mL,用于测定面粉、血清、中药等几种样品中的痕量铜,结果满意。相对标准偏差为2.0%~2.9%。 new catalytic kinetic spectrophotometricmethod for the determination of trace copper(Ⅱ)based on catalytic effect of Cu(Ⅱ)on the decolorizingreaction resulted from the reduction ofdichlofobromoarsenazo with ascorbic acid has beeninvestigated.The order of the reaction and theapparent activation energy have been determined andthe reaction mechanism has been discussed.Thedetection limit is 7.6 ×10 ̄(-12)g/mL and linear range is0~60ng/10mL. The method has been applied to thedetermination of copper in samples with satisfactoryresults and the relative standard deviation is 2.0~2.9%。
出处 《分析试验室》 CAS CSCD 北大核心 1994年第6期42-45,共4页 Chinese Journal of Analysis Laboratory
关键词 催化光度法 面粉 血清 Catalytic kinetic spectrophotometry dichlorobromoarsenazo copper determination
  • 相关文献

参考文献1

二级参考文献4

  • 1徐友宜,北京大学学报,1992年,28卷,200页
  • 2李南强,化学学报,1984年,42卷,1057页
  • 3Cheng K L,CRC Handbook of Organic Analytical Reagents,1982年
  • 4李南强,分析化学,1973年,1卷,40页

共引文献17

同被引文献270

引证文献27

二级引证文献124

相关作者