期刊文献+

毒饵及生物材料中F^-的选择电极法测定 被引量:1

Fluoride Detecting in Biomaterial and Toxic Bait by Ion-selective Electrode
全文增补中
导出
摘要 笔者采用选择电极电位法,测定毒饵及生物材料中F^-含量,方法简便快速,结果可信,相对偏差<5%。线性范围在0.019~1900mg/L。主要干扰Al^(3+)及影响测定体系的诸因素,选用总离子强度调节缓冲液(TISAB)掩蔽调节作用,测得结果满意。 This article presents the study of detecting fluoride(F^-) in toxic bait and biomaterial using selective potentiomatirie electrode method.The method had not only quick, but also fine selectivity and stability. The linear range was 0.019-1900mg/L, and the relative deviation was no more than 5%.It interfered Al^(3+) and effects various of determination system princepally.The method used masking and adjustment of total ionic strength adjustment buffer (TISAB)reducing interfering Al^(3+) and other factors got good results.
出处 《中国媒介生物学及控制杂志》 CAS CSCD 1993年第6期434-436,共3页 Chinese Journal of Vector Biology and Control
关键词 毒饵 生物材料 氟离子 灭鼠 Toxic bait Biomateria Fluoride TISAB