摘要 国产200mA X线机不仅不能摄高电压胸片,而且是产生废片的主要机型。估计全国每年所摄尘肺胸片中有60万张以上废片。本研究设想通过工艺改造,将此型X线机改为输出125kV用以摄尘肺高电压胸片。可望大幅度提高一级片率,降低废片率。本研究通过重组高压系统及控制电路达到目的。现场试验结果示:摄高电压胸片结果满意,共摄片1707张,获一级片分别为45.7%和52.2%,废片率0.9%和2.3%。改造200mA X线机摄高电压尘肺胸片是现阶段的一条经济、实用解决现有低档机、提高胸片质量的方法,值得推广。 Rdiograph is the first modality for pneumoniesis surveillance, but an ordinary. X-rayunit with capacity of 200 mA 100kV can notfulfill this task. It is roughly estimated thatmore than 0.6 million unqulified chest radio-graphs produced each years with those unit.Reconstruction of those unit make them moreeffective to achieve high kilovoltage techniqueand consequently improve the quality of chestfilms. 1707 films were taken by 2 new units,the study showed that the excellent films ratewere 42.7% and 52.2% respectivity and filmunreadable were less then 3%.
出处 《中国工业医学杂志》 CAS 1993年第4期193-197,共5页 Chinese Journal of Industrial Medicine
关键词 尘肺 胸片 X线机 诊断 pneumoconiosis chest radiograph quality assurance
  • 相关文献

同被引文献23

  • 1丁茂柏.尘肺综合诊断指标的研究[J].中国工业医学杂志,1991,4(2):8-11. 被引量:4
  • 2孙承业,丁茂柏,范焱,张翠娟,刘平生.尘肺高、低电压X线摄影技术对比研究[J].中华劳动卫生职业病杂志,1994,12(3):4-6. 被引量:5
  • 3郭鼐,田琳,张慧峰.肺灌洗治疗尘肺的现状与刍见[J]中华劳动卫生职业病杂志,1994(04).
  • 4胡秀云,张林清,苏光跃,杨少泉.煤矿尘肺的流行病学分析[J]中华劳动卫生职业病杂志,1986(01).
  • 5郑荫柯,王保林,童育贤,苗素爱.一个掘进队湿式凿岩20年防尘效果评价[J]中华劳动卫生职业病杂志,1985(06).
  • 6楼介治.调离矽尘作业工人矽肺发病情况的调查[J]中华劳动卫生职业病杂志,1984(06).
  • 7吴执中,刘荫曾,张翠娟,陈珍,周晓蓉,张辉明.某钨矿矽肺发病情况的回顾性调查[J]中华劳动卫生职业病杂志,1983(01).
  • 8鲁锡荣,王亿诚,张煌保,王英奎.三万分子量克矽平治疗矽肺病三年小结[J]卫生研究,1982(01).
  • 9吴执中,于得汶.我国尘肺防治与研究的回顾和前瞻(综述)[J]卫生研究,1980(04).
  • 10吴执中.尘(矽)肺的发病情况、诊断及治疗问题[J]卫生研究,1979(Z1).