摘要 The data of uranium isotope abundance ratios measured by thermal ionization mass spectrometer MAT-261 were analyzed in the three aspects including distribution function, correlation and uncertainty. The measurement data were confirmed to be normal distribution. Uncertainty and correlation of the isotope abundance ratios measured by MAT-261 in the certified uranium sample were investigated. The conclusion obtained indicated that, the correlation only between abundance ratios R5/8 and R4/8 is strong, but the others is weak for the enriched uranium sample; For the natural uranium sample, the correlation among all abundance ratios is weak. On this base, the uncertainty of all abundance ratios measured by MAT-261 had been evaluated.
出处 《质谱学报》 EI CAS CSCD 2004年第B10期187-188,共2页 Journal of Chinese Mass Spectrometry Society
  • 相关文献

参考文献2

  • 1朱凤蓉.UF6标准物质中铀同位素丰度的质谱测量[R].西安:西北核技术研究所,..
  • 2C利特安努 L 里卡著.数理统计理论与方法在痕量分析中的应用[M].北京:原子能出版社,1990.41.