摘要 目的 通过将复合有重组人类骨形态发生蛋白 2 (rhBMP 2 )的注射型磷酸钙人工骨 (rhBMP 2 /CPC)应用于体外经皮椎体成形术 (PVP)实验中 ,观察并评价其生物力学性能 ,以探寻rhBMP 2 /CPC能否有效替代聚甲基丙烯酸甲酯 (PMMA)而应用于PVP手术。方法 选取 6具尸体脊柱标本 (T8~L5)并测试骨密度 ,然后将其游离成 60个椎体并随机分为 4组 :I组为空白对照组 (n =3 0 ) ;II组为PMMA组 (n =10 ) ;III组为CPC组 (n =10 ) ;IV组为rhBMP 2 /CPC组 (n =10 )。在“C”形臂X线机导引下 ,II~IV组椎体经双侧椎弓根分别充填PMMA、CPC及rhBMP 2 /CPC ;在万能材料实验机上将所有椎体压缩至原始高度的 5 0 % ,测得最大载荷及刚度 ;随后将I组再进一步划分为IA、IB及IC组 ,分别充填PMMA、CPC及rhBMP 2 /CPC ,按前述加载方法再次压缩至充填后椎体高度的 5 0 %。结果 实验开始前 ,各组椎体BMD为 ( 0 .3 10~ 0 .689)g/cm2 ,高度为 ( 1.64~ 3 .0 9)cm (组间比较 ,均P >0 .0 5 )。IA、IB及IC组充填后椎体前缘高度分别增加 ( 0 .16±0 .0 5 )cm ,( 0 .18± 0 .0 7)cm和 ( 0 .2 1± 0 .0 9)cm ,与充填前比较 ,差异无显著性意义 (均P >0 .0 5 )。材料充填量方面 :IA组是 ( 2 .8± 0 .7)ml,IB组是 ( 2 .3± 0 .3 )ml,IC组是 ( Objective To evaluate the biomechanical performance of the injectable calcium phosphate cement (CPC) recombined human bone morphogenetic protein-2(rhBMP-2/CPC) in an in vitro vertebroplasty model. Methods A total of 60 vertebral bodies(VBs) obtained from 6 cadaveric thoracolumbar spine (T 8~L 5) were used.After the bone mineral density (BMD) of each VB was measured using a dual energy X-ray absorptiometry (DEXA),the VBs were assigned into 4 groups,intact control group (group I),polymethylmethacrylate (PMMA) injection group(group II),CPC injection group (group III) and rhBMP-2/CPC injection group(group IV). The VBs were then fixed on a material testing system and compressed to 50% of their original heights,respectively. After fracture,the VBs in group I were further divided into 3 subgroups and injected with PMMA(group IA),CPC(group IB) or rhBMP-2/CPC(group IC),respectively. The VB height was measured again after cement injection to evaluate the magnitude of the regained height,and these VBs were compressed to 50% of the post-injection height. Force and displacement data were recorded and used to determine the maximum load and stiffness. Results BMDs of all tested groups ranged from 0.310 to 0.689 g/cm 2( P >0.05),VBs height ranged from 1.64 to 3.09 cm( P >0.05). For groups II,III,IV,IA,IB and IC,3.2±2.6ml,3.1±1.9ml,2.9±1.4ml,2.8±0.7ml,2.3±0.3ml and 2.4±0.3 ml of cement were injected on average( P >0.05),respectively. No significant difference was revealed in the mean VB height regained by injection of PMMA. The maximum load of groups II,III,IV and IA were greater than that of group I ( P <0.05) ,no significant difference was found among groups IC,IB and I. Stiffness of groups II,III,IV and IA were greater than that of group I( P <0.05),and the stiffness of group I was greater than that of group IC( P <0.05),while no significant difference was found between groups I and IB( P >0.05). Conclusion CPC and rhBMP-2/CPC can improve the biomechanical strength of the VB,and could be used as the substitute of the PMMA in the percutaneous vertebroplasty operation.
出处 《中华物理医学与康复杂志》 CAS CSCD 北大核心 2004年第7期392-395,共4页 Chinese Journal of Physical Medicine and Rehabilitation
关键词 复合RHBMP-2 注射型磷酸钙 人工骨 椎体成形术 生物力学 充填材料 Osteoporosis Biomechanics Vertebroplasty PMMA CPC rhBMP-2 rhBMP-2/CPC
  • 相关文献

参考文献17

  • 1Deramond H,Mathis JM.Vertebroplasty in osteoporosis.Semin Musculoskel Radiol,2002,6:263-268.
  • 2Galibert P,Deramond H,Rosat P, et al.Preliminary note on the treatment of vertebral angioma by percutaneous acrylic vertebroplasty. Neurochirurgie, 1987, 33:166-168.
  • 3Deramond H,Galibert P,Debussche C.Vertebroplasty.Neuroradiology,1991,33:177-178.
  • 4Perez-Higueras A,Alvarez L,Rossi RE,et al.Percutaneous vertebroplasty:long-term clinical and radiological outcome.Neuroradiology,2002,44:950-954.
  • 5Watts NB,Harris ST,Genant HK.Treatment of painful osteoporotic vertebral fractures with percutaneous vertebroplasty or kyphoplasty.Osteoporos Int,2001,12:429-437.
  • 6Grados F,Depriester C,Cayrolle G,et al.Long-term observation of vertebral osteoporotic fractures treated by percutaneous vertebroplasty.Rheumatology,2000,39:1410-1414.
  • 7Evans AJ,Jensen ME,Kip KE,et al.Vertebral compression fracture:pain reduction and improvement in functional mobility after percutaneous polymethylmethacrylate vertebroplasty retrospective report of 245 cases.Radiology,2003,226:366-372.
  • 8Berlemann U,Ferguson SJ,Nolte LP,et al.Adjacent vertebral failure after vertebroplasty:a biomechanical investigation.J Bone Joint Surg,2002,84B:748-752.
  • 9Garfin SR,Yuan HA,Reiley MA.New technologies in spine:kyphoplasty and vertebroplasty for the treatment of painful osteoporotic compression fractures.Spine,2001,26:1511-1515.
  • 10Lim TH,Brebach GT,Renner SM,et al.Biomechanical evaluation of an injectable calcium phosphate cement for vertebroplasty.Spine,2002,27:1297-1302.