单层螺旋CT最佳图像密度分辨率时扫描参数的探讨 被引量:2

Research and Discussion of Scanning Parameters of SingleHelical CT Best Density Resolution
暂未订购
导出
摘要 目的 :分析不同扫描参数对图像质量及密度分辨率的影响 ,探讨单层螺旋CT在最佳密度分辨率时的扫描参数。结果 :当kV不变时 ,随着mAs的增加 ,图像的噪声得到改善 ,密度分辨率也相应提高。同时密度分辨率随着层厚、重建视野的增大而提高 ;随着扫描物体厚度、重建矩阵、螺距增大而降低 ;平滑重建算法可以提高密度分辨率。在最佳图像密度分辨率时的扫描参数为 :10 0kV与 375mAs、 12 0kV与 30 0mAs、14 0kV与 30 0mAs、 10mm层厚、 2 4 0mm重建视野 ,平滑重建算法。在最佳图像密度分辨率时的扫描参数应用到临床肝脏扫描 ,能提高图像质量、提高密度分辨率 ,对临床具有十分重要的意义。 Objective:To determine the optional scanning parameters used in single helical CT Scanning on best density resolution by analyzing the effect of different scanning parameters on the image quality and density resolution.Results:where the kV was fixed,the image noise was decreasing with the increasing of mAs,white density resolution was increased,too.when mAs was fixed,the same results occurred with the increase of kV.Density resolution was increased with increasing of slice thickness.FOV,and decreased with matrix,pitch and thickness of scanned objective.The smooth reconstruction results to increase density resolution.Scanning parameters on best density resolution were:120?kV and 375?mAs,120?kV and 300?mAs,140?kV and 300?mAs,10?mm slice thickness 240?mm FOV,smooth reconstruction.Conclusion scanning parameters on best density resolution use to scan livers,It increases the image quality and density resolution.It is very important in clinic.
出处 《医疗装备》 2004年第8期1-3,共3页 Medical Equipment
关键词 单层螺旋CT 最佳图像 密度分辨率 扫描参数 图像质量 重建算法 Computed Tomography Density Resolution Noise Scanning parameters
  • 相关文献

参考文献4

  • 1黄大同.CT扫描机的验收检测[J].中华放射学杂志,1994,28(9):616-619. 被引量:13
  • 2IEC 1223-2-6 Constancy test X-ray equipment for com puted tomography international electro technical commission 1994,1.
  • 3Crespi A,Leoni S,Morotanari G,et al; The evaluation of the physical characteristics of a volumetric computer tomography.Radiol Med Ctorino.1996,91:460.
  • 4GB/T17589-1998.X射线计算机断层摄影装置影像质量保证检测规范:中华人民共和国国家标准.[S].(北京)国家质量技术监督局,卫生部,1999,1..

共引文献12

同被引文献32

  • 1柳澄,秦维昌.多层螺旋CT(一)[J].医学影像学杂志,2000,10(3):194-195. 被引量:42
  • 2杨裕华.X线CT影像技术研究进展[J].医疗设备信息,2004,19(9):25-28. 被引量:12
  • 3MATHEWS J D, FORSYTHE A V, BRADY Z, et al. Cancer risk in 680, 000 people exposed to computed tomography scans in childhood or adolescence : data linkage study of 11 million Austra- lians[J]. BMJ, 2013, 346(7910): 1-18.
  • 4BENZ M G, BENZ M W, BIRNBAUM S B, et al. Improved spa- tial resolution and lower- dose pediatric CT imaging: a feasibility study to evaluate narrowing the X - ray photon energy spectrum [J]. Pediatr RadioL, 2014, 44(8) : 1026 - 1030.
  • 5RIZZO S, KALRA M, SCHMIDT B, et al. Comparison of angular and combined automatic tube current modulation techniques with constant tube current CT of the abdomen and pelvis[ Jl. AJR Am J Roentgenol, 2006, 186 (3) : 673 - 679.
  • 6STRAUSS K J, GOSKE M J, FRUSH D P, et al. Image Gently Vendor Summit: working together for better estimates of pediatric radiation dose from CT[ J]. AJR Am J Roentgenol, 2009, 192 (5) : 1169 -1175.
  • 7LEE S M, LEE W, CHUNG J W, et al. Effect of kVp on image quality and accuracy in corona~ CT angiography according to pa- tient body size: a phantom study [ J ]. Int J Cardiovasc Imaging, 2013, 29(2) : 83 -9l.
  • 8MOSCAR]ELLO A, TAKX R A, SCHOEPF U J, et al. Coronary CT angiography: image quality, diagnostic accuracy, and potential for radiation dose reduction using a novel iterative image recon- struction technique - comparison with traditional filtered back pro- iection[J]. Eur Radiol, 2011, 21(10) : 2130 -2138.
  • 9SONG J S, LEE J M, SOHN J Y, et al. Hybrid iterative recon- struction technique for liver CT scans for image noise reduction and image quality improvement: evaluation of the optimal iterative re- construction strengths[ J]. Radiol Med, 2014, 120 (3) : 259 - 267.
  • 10BAUMUELLER S, WINKLEHNER A, KARLO C, et al. Low - dose CT of the lung: potential value of iterative reconstructions [J]. EurRadiol, 2012, 22(12) : 2597 -2606.

引证文献2

二级引证文献5

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部