摘要 目的 观察使用Symmetry主动脉吻合器时对大隐静脉移植物内皮有无损伤。 方法 对 10例使用Symmetry主动脉吻合器行冠状动脉旁路移植术患者进行自身对照研究。在主动脉吻合器使用前和使用后 ,切取自体大隐静脉片断 ,进行Ⅷ因子免疫组化染色 ,对照血管内皮损伤情况。结果 血管内皮细胞均匀着色 ,为深褐色颗粒 ,损伤处褐色颗粒排列中断。吻合器组内皮损伤处平均为 2 .5 5处 ,对照组为 2 .0处。 Purpose: To evaluate the injury with Symmetry aortic connector to saphenous venous endothelial cells. Methods: Segments of vein graft from 10 patients received coronary artery bypass grafting with Symmetry aortic connector were stained histologically with hematoxylin eosin and factor VIII immunochemical stain. Endothelial cells injury was examined and counted. Results: Average injury site in Symmetry aortic connector group and the control group was 2.55 and 2.0 respectively (P < 0.05). Conclusions: Symmetry aortic connector does not increase the injury to venous endothelial cells.
出处 《复旦学报(医学版)》 EI CAS CSCD 北大核心 2004年第4期401-402,共2页 Fudan University Journal of Medical Sciences
关键词 移植物 Symmetry主动脉吻合器 大隐静脉 内皮细胞 Blood vessels Grafts Histology Medical problems Patient monitoring
  • 相关文献

参考文献6

  • 1Eckstein FS,Bonilla LF,Englberger L,et la.Minimizing aortic manipulation during OPCAB using the symmetry aortic connector system for proximal vein graft anastomoses.Ann Thorac Surg,2001,72:S995
  • 2Carrel TP,Eckstein FS,Englberger L,et la.Pitfalls and key lessons with the symmetry proximal anastomotic device in coronary artery bypass surgery.Ann Thorac Surg,2003,75:1434
  • 3Grondin CM,Campeau L,Lesperance J,et al.Comparison of late changes in internal mammary artery and saphenous vein grafts in two consecutive series of patients 10 years after operation.Circulation,1984,70:I208
  • 4Antona C,Scrofani R,Lemma M,et al.Assessment of an aortosaphenous vein graft anastomotic device in coronary surgery:clinical experience and early angiographic results.Ann Thorac Surg,2002,74:2101
  • 5Wiklund L,Bugge M,Berglin E.Angiographic results after the use of a sutureless aortic connector for proximal vein graft anastomoses.Ann Thorac Surg,2002,73:1993
  • 6Verma S,Fedak P,Ko L,et al.Evaluation of a novel sutureless anastomotic connector:from endothelial function to mid-term clinical and angiographic follow-up.J Thorac Cardiovasc Surg,2003,126:1555