期刊文献+

薄膜和纤维塑料闪烁体中子能响特性的Monte Carlo数值计算 被引量:7

Monte Carlo modeling of neutron sensitivity for foil- or fiber- plastic scintillator
在线阅读 下载PDF
导出
摘要 根据中子与塑料闪烁体的作用机制及塑料闪烁体的荧光转换特性,研究了塑料闪烁体薄膜和塑料闪烁体纤维中子能响特性的MonteCarlo数值计算方法,给出了不同厚度的塑料闪烁体薄膜和不同直径的塑料闪烁体纤维对裂变能区中子的灵敏度计算结果,并对计算结果及其在电流型坪响应裂变中子探测器方面可能的应用进行了初步探讨。 Based on the key reaction with neutrons within plastic scintillator and its fluorescence characteristic, Monte Carlo simulation method is introduced for neutron sensitivity calculation for foil- or fiber- plastic scintillator. The energy dependent sensitivities in the fission neutron energy range are calculated and discussed for various foil thickness and fiber diameters, which suggest a possible flat-response current-mode detector scheme for fission neutron measurement through combination of plastic scintillator foils with different thickness.
出处 《核电子学与探测技术》 CAS CSCD 北大核心 2004年第4期346-349,334,共5页 Nuclear Electronics & Detection Technology
基金 中国工程物理研究院基金(20030209) 惯性约束聚变技术探索基金(2003AA84ts20)资助的项目
关键词 塑料闪烁体 中子灵敏度 蒙特卡罗模拟 中子输运 plastic scintillator neutron sensitivity Monte Carlo method
  • 相关文献

参考文献9

  • 1[1]欧阳晓平.低强度脉冲裂变中子探测技术研究[M].上海:复旦大学,2002.
  • 2赵有雄,詹文龙,郭忠言,郗洪飞,周建群,雷怀宏,王金川,罗永峰,冯恩普.椭球面聚焦型镜像塑料闪烁探测器及其应用[J].核电子学与探测技术,1995,15(5):270-274. 被引量:2
  • 3[3]Ress D, et al. High- sensitivity scintillating-fiber imaging detector for high-energy neutrons[J]. Rev Sci Instrum,1995, 66:4943.
  • 4[4]Morgan GL, et al. Development of a neutron imaging diagnostic for inertial confinement fusion experiments[J]. Rev. Sci. Instrum,2001, 72:865.
  • 5[5]张传飞.脉冲裂变中子总数测量新方法研究[D].成都:四川大学,2002.
  • 6[7]Fishman GS. Monte Carlo-Concepts, Algorithms and Applications[M]. Springer, New York, 1996.
  • 7[8]Briesmeister JF. MCNP-A General Monte Carlo N-Particle Transport Code[R]. LA-12625-M, Version 4B, 1997.
  • 8[9]Cullen DE. TART98 a coupled neutron-photon 3-D, combinatorial geometry time dependent monte carlo transport code[R].UCRL-ID- 126455-REV-2.
  • 9[10]Brochure of Saint-Gobain Scintillation Products. Organic Scintillators.

二级参考文献9

  • 1Galindo-Uribarri A,,Drakc TE,Ball GC et al.A thin plastic scintillator time-zero detector and flux monitor. Nuclear Instruments and Methods . 1991
  • 2Buenerd M,Ilendrie DL,Jahnke U et al.Response of pilot u scintillator to heavy ions. Nuclear Instruments and Methods . 1976
  • 3Muga ML,Burnsed DJ,Steeger WE et al.A new time of flight particle detector. Nuclear Instruments and Methods . 1970
  • 4Philips Data Handbook Photomultipiers PC04. . 1990
  • 5Lolos GJ,Iluber GM,Ramadan B et al.An improved amplification resistor chain for photomultiplier tubes. Nuclear Instruments and Methods . 1991
  • 6Zhan Wenlog,Guo Zhongyan,Zhou Jianqun et al.Radioactive nuclear beam line at HIRFL. Physics&Nuclear Phsycs . 1994
  • 7Becchetti FD,Thorn CE,Levine MJ.Response of plastic scintillator detectors to heavy ions,Z≤35,E≤170 MeV. Nuclear Instruments and Methods . 1976
  • 8T. Batsch and M. Moszyski.Timing properties of thin scintillator foils. Nuclear Instruments . 1975
  • 9C. K. Gelbke,K. D. Hildenbrand and R. Bock.A time-of-flight spectrometer for heavy ions. Nuclear Instruments . 1971

共引文献1

同被引文献58

引证文献7

二级引证文献27