Precil XC-20型自动血沉仪的应用评价与血沉测定的方法进展

Methological Dscussion of ESR Test and Application Evaluation for Precil XC-20 Automated ESR Analyzer
在线阅读 下载PDF
导出
摘要 目的 :比较PrecilXC - 2 0型自动血沉仪测定结果与传统魏氏法的相关性和线性。方法 :分别用自动血沉仪法和传统魏氏法测定 5 0例临床病人 1h血沉。结果 :经统计学处理 ,两种方法的测定结果相关系数 0 .84 ,在正常参考值范围内 ,二者测定结果一致 ,但当血沉 >10mm/h后 ,手工法测定均值高于仪器法 ,P <0 .0 1,而且随着测定值增加 ,二者之间差异越明显。结论 :自动血沉仪法对正常标本测定结果与手工法接近 。 Objective: To compare the corralation and linerization of erythrocyte sedimentation rate(ESR)between the automated ESR analyzer and manual Westergren method. Methods: test the clinical specim by means of automated ESR analyzer(Precil XC-20) and manual Westergren method for 50 cases separately. Results: The corralation coeffecion was 0.84,and there was no significant difference of the ESR value between the two method in the normal range(ESR<15mm/h,P>0.05),while the difference was very significant when the ESR > 15mm/h(P<0.01).And the results of manual Westergren were higher than the automated method.Conclusion: The results were similar between the two methods in that of normal walues with ESR,and it was more reliable by the manual Westergren method when the ESR >15mm/h.
出处 《张家口医学院学报》 2004年第3期1-3,共3页
关键词 PrecilXC-20型自动血沉仪 血沉测定 Westergren法 血沉管 erythrocyte sedimentation rate(ESR) detection for ESR automated ESR analyzer Westergren method
  • 相关文献

参考文献6

  • 1王海,傅淑宏,丛玉隆.红细胞沉降率动态图形对临床诊断意义的探讨[J].临床检验杂志,1999,17(5):294-295. 被引量:6
  • 2NCCLS. Renference and Selected procedure for the ErythrocyteSedimentation Rate ( ESR ) Test. Approved Standard - FourEdition. 2000, 20(27): 1 - 10.
  • 3Mario Plebani, MD, Stefania De Toni, MD, Maria Colombma Sanzari, MD. The Testl Automated System, A New Method for Measuring the Erythrocyte Sedimatation Rate. et al. Am J Clin Pathol, 1998, 110(3) :334 - 340
  • 4Davide Giaarina; Lilbano Capuzzo; Fabrizio Cauduro; et al. Internal quality contarol for erythrocyte sedimentation rate measured by test - 1 analyzer. Clin Lab: 2002, 48(9/10): 459-462
  • 5Plebani, Mario; Piva, Elisa. Erythrocyte sedimentation rate:use of fresh blood for quality control. Am J Clin Pathol, 2002,117(4):621 - 626
  • 6黄玉霞,唐书强,杨伟男,潘万德,刘际芬,王仲琼,周蕴苓,彭黎明.Monitor-J型自动血沉仪分析结果初探[J].华西医学,1999,14(2):236-237. 被引量:4