摘要 目的 对电化学发光免疫法与微粒子酶联免疫法检测血清甲胎蛋白作对比分析。方法 随机抽取150份临床送检的标本,用电化学发光免疫法与微粒子酶联免疫法测试甲胎蛋白的含量,对测试数据进行相关性试验、线性试验、精密度试验、回收试验。结果 两法无显著性差异(P>0.05),但电化学发光免疫法的线性范围较宽(1-980ng/ml),重复性较好(批内CV值3.4%,批间CV值3.5%),回收率较高(97.2%)。结论 Objective To evaluate the efficiencies of electrochemiluminescent assay (ECL) andmicroparticle enzyme immunoassy(MEIA)for detecting of the levels of serum AFP .Methods Serum AFPlevels of 150 samples were determined with both ECL and MEIA. The correlation between ECL and MEIA wereassessed, and the assays of linear range, recovery and precision of the two methods were performed andalso evaluated.Results There was no significant difference between ECL and MEIA for detection of thelevels of serum AFP (P>0.05). Overall, however, in comparison with MEIA, ECL had a broader linear range(1-980ng/ml) and a higher precision (the within-run, 3.4%; the between-run, 3.5%) as well as a higheraverage recovery rate (97.2%).Conclusions There was a good correlation between ECL and MEIA fordetermination of the levels of serum AFP. In general, for detecting the levels of serum AFP, ECL wassuperior to MEIA.
出处 《国际医药卫生导报》 2003年第22期9-10,共2页 International Medicine and Health Guidance News
  • 相关文献

参考文献1

二级参考文献9

  • 1韩松.快速斑点免疫结合试验的进展[J].中华医学检验杂志,1994,17(4):249-250. 被引量:15
  • 2陶义训.快速斑点免疫结合试验在医学检验中的应用评价[J].中华医学检验杂志,1994,17:197-197.
  • 3俞海燕,临床检验杂志,1998年,5卷,118页
  • 4郑佐娅,临床检验杂志,1998年,5卷,10页
  • 5陶义训,免疫学和免疫学检验(第2版),1997年,140页
  • 6韩松,中华医学检验杂志,1994年,17卷,249页
  • 7陶义训,中华医学检验杂志,1994年,17卷,197页
  • 8郑佐娅,陶义训.电化学发光免疫测定[J].临床检验信息导报,1998,5(1):10-12. 被引量:15
  • 9俞海燕,陶义训.免疫层析试验的临床应用进展[J].临床检验信息导报,1998,5(4):118-120. 被引量:6

共引文献31

同被引文献17

引证文献2

二级引证文献3

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部