摘要 在医学图像导航微创外科手术中,系统定位是一个非常关键的问题。为了在医疗微创手术中利用磁定位器定位,对磁定位器的精度进行了测定。首先介绍了国外在磁定位器精度测试方面的研究现状,然后设计了一个磁定位器精度分析的实验平台,并在此平台上进行了磁定位器精度性能评价的实验,通过实验数据的分析处理,得出相关结论。实验结果表明,磁定位器的自身精度高,能够满足医疗手术中应用的要求;同时,医疗环境的合理布置,可以避免磁定位器受到干扰,从而进一步验证了磁定位器在医疗手术中应用的可行性。 In surgical image-guided minimally invasive surgery (MIS),it s a key issue to determinate the location of the system accurately. The major aim of this paper is to evaluate the accuracy of electromagnetic tracking devices (EMTD) for surgical use. It first presents the current worldwide research situation on the accuracy analysis of EMTD, and then designs a framework to analyze the accuracy of EMTD. Some experiments for assessing the performance of EMTD have been completed on the basis of this framework. Resu...
出处 《机械设计与制造》 北大核心 2008年第2期92-94,共3页 Machinery Design & Manufacture
基金 国家自然科学基金项目(编号:50675109) 北京市科委资助项目(编号:H060720050330)
关键词 磁定位 医疗环境 位姿 精度分析 Electromagnetic tracking Surgical environment Position and pose Accuracy analysis
  • 相关文献

参考文献2

  • 1[1]Wu.X.H.,Taylor,R..A direction space interpolation technique for cali bration of electro-magnetic surgical navigation systems[C]//MICCAI 2003:Lecture Notes in Computer Science(2879),Bedin:Springer-Vedag,2003:215-222
  • 2[2]Emmanuel Wilson.Accuracy Analysis of Electromagnetic Tracking within Medical Environments[D].Washingtorr Georgetown University,2006

同被引文献21

  • 1杨柳,陈艳萍.一种新的Levenberg-Marquardt算法的收敛性[J].计算数学,2005,27(1):55-62. 被引量:41
  • 2谢水生,李华,林宗楷,王少荣,陈增淦,张键,陈统一.基于连续组织切片的神经三维可视化研究[J].计算机应用研究,2007,24(6):170-172. 被引量:9
  • 3Roberts DW,Steohbehn JW,Hatch JF,et al. frameless stereotaxic integra tion of computerized tomographic imaging and the operating microscope [ J ] .N eurosurg, 1986( 65 ) : 546--549.
  • 4Davies,S.Starkie,SJ.Harris,ct al.Neurobot:A special-purpose robot for Neurosurgery[ C ].Proc.Of 2000 IEEE Int.Conf.on Robotics and Automation.2000:4103--4108.
  • 5F.Tatar,J.R.Mollingcr, R.C.Den Dulk,et al.Ultrasonic sensor system for measuring position and orientation of laproseopic instruments in minimal invasive surgery [ C ].2nd Annual Internation IEEE-EMB Special Topic Conference on Microtechnologies in Medicine&Biology, 2002: 301-304.
  • 6R.L.Galloway,R.J macinuna,W.A.Bass,el aLAn optical device for interac tire,image-guided neurosurgery [C].Proc.Of the 15th IEEE Annual International Conference on Engineering in medicine Biology Society, 1993 ( 10 ) : 28-31,954-955.
  • 7t S.A.Tebo,D.A.Leopold,D.M&ong,et alan optical 3D digitizer for frameless stereotactlc surgery [ J ].IEEE computer Graphics and Applications, Jan. 1996:55-64.
  • 8Boctor E M, Fischer G, Choti M A,et al.A Dual-Armed Robotic System for Intraoperative Ultrasound Guided Hepatic Ablative Therapy:a Prospective Study [ C ].IEEE.Proeeedings of the 2004 IEEE International Conference on Robotics and Automation. New Orleans,LA: IEEE ,2004: 2517-2522.
  • 9徐静.基于超声影像导航的肝癌消融机器人系统精度研究[D].北京:清华大学出版社,2007.
  • 10D.D.Frantz,A.D.Wiles,S.E.Leis et al.Accuracy assessment protocols for electromagnetic tracking systems[J].Physics in medicine and biology,2003,48(14):2241-2251.