This study reports the performances of efficiency calibrations for high-purity-germanium gamma-ray spectrometry using the source-, Laboratory Sourceless Object Calibration Software(LabSOCS)-and ANGLE-based methods in ...This study reports the performances of efficiency calibrations for high-purity-germanium gamma-ray spectrometry using the source-, Laboratory Sourceless Object Calibration Software(LabSOCS)-and ANGLE-based methods in an inter-comparison exercise. Although the results of LabSOCS and ANGLE for ^(241)Am emitting lowenergy gamma rays were not very satisfactory, all of the three efficiency calibration methods passed acceptance criteria. The results confirmed the reliability of the calculation codes ANGLE and LabSOCS as alternative efficiency calibration methods in high-purity-germanium gamma spectrometry. This study is likely to promote the further application of the ANGLE and LabSOCS calculation codes in radioactivity measurements.展开更多
基金supported by the Support Program of the Ministry of Science and Technology(No.2014FY211000)the National Key Technology Research and Development Program(No.2013BAK03B05)
文摘This study reports the performances of efficiency calibrations for high-purity-germanium gamma-ray spectrometry using the source-, Laboratory Sourceless Object Calibration Software(LabSOCS)-and ANGLE-based methods in an inter-comparison exercise. Although the results of LabSOCS and ANGLE for ^(241)Am emitting lowenergy gamma rays were not very satisfactory, all of the three efficiency calibration methods passed acceptance criteria. The results confirmed the reliability of the calculation codes ANGLE and LabSOCS as alternative efficiency calibration methods in high-purity-germanium gamma spectrometry. This study is likely to promote the further application of the ANGLE and LabSOCS calculation codes in radioactivity measurements.