AIM:To study the relationship between ultrasound biomicroscopy(UBM)and Lenstar when measuring anterior chamber depth(ACD)in eyes with shallow anterior chamber,and the necessity of conducting both examinations.METHODS:...AIM:To study the relationship between ultrasound biomicroscopy(UBM)and Lenstar when measuring anterior chamber depth(ACD)in eyes with shallow anterior chamber,and the necessity of conducting both examinations.METHODS:This is a retrospective observational study including 56 acute primary angle closure(APAC)eyes and 47 primary angle closure suspect(PACS)eyes with shallow ACD.ACD value measured by Lenstar and UBM were documented.The Bland-Altman plots were examined separately in all included eyes,APAC eyes and PACS eyes,for the assessment of agreement between two measurements.The agreement was compared across different population by evaluating mean difference,width of 95%limit of agreement(LoA)and the presence of proportional bias or outliers in Bland-Altman plots.R E S U LT S:The average ACD in APAC eyes(1.71±0.23 mm)was significantly smaller than that in PACS eyes(1.79±0.25 mm,P=0.038).Bland-Altman plots of both APAC eyes and PACS eyes showed small mean difference without the presence of proportional bias.However,compared with PACS eyes,the Bland-Altman plot of APAC eyes had wider 95%LoA and more outliers outside the 95%LoA.CONCLUSION:Despite the small ACD values,the two methods in measuring ACD,Lenstar and UBM,demonstrate substantial consistency when measuring PACS eyes.However,for the APAC population,the agreement might be limited,and thus the measured values should be interpreted with caution.展开更多
目的比较lenstar LS 900与角膜超声测厚仪测量人中央角膜厚度(CCT)的差异,为临床检查提供参考。方法选取屈光手术者26例(52只眼),分别用lenstar LS 900和角膜超声测厚仪测量双眼CCT,两种仪器均测量3次或以上,取其平均值,对所得数据进行...目的比较lenstar LS 900与角膜超声测厚仪测量人中央角膜厚度(CCT)的差异,为临床检查提供参考。方法选取屈光手术者26例(52只眼),分别用lenstar LS 900和角膜超声测厚仪测量双眼CCT,两种仪器均测量3次或以上,取其平均值,对所得数据进行统计学分析。结果 52只眼CCT测量结果:lenstar LS 900为(538.27±40.38)μm,A超为(534.15±41.12)μm。两种测量方法结果差异有统计学意义(t=2.9643,P<0.005)。但测量结果高度相关(r=0.970),前者所测CCT较后者厚(4.11±10.11)μm。结论 lenstar LS 900与角膜超声测厚仪对CCT的测量结果呈较高的相关性,两种仪器具有较高的测量准确性,lenstar LS 900可以作为角膜厚度的测量工具之一。展开更多
基金Supported by the Surface Project of National Natural Science Foundation of China(No.82070957)the Shandong Medical and Health Technology Development Project(No.202107021077).
文摘AIM:To study the relationship between ultrasound biomicroscopy(UBM)and Lenstar when measuring anterior chamber depth(ACD)in eyes with shallow anterior chamber,and the necessity of conducting both examinations.METHODS:This is a retrospective observational study including 56 acute primary angle closure(APAC)eyes and 47 primary angle closure suspect(PACS)eyes with shallow ACD.ACD value measured by Lenstar and UBM were documented.The Bland-Altman plots were examined separately in all included eyes,APAC eyes and PACS eyes,for the assessment of agreement between two measurements.The agreement was compared across different population by evaluating mean difference,width of 95%limit of agreement(LoA)and the presence of proportional bias or outliers in Bland-Altman plots.R E S U LT S:The average ACD in APAC eyes(1.71±0.23 mm)was significantly smaller than that in PACS eyes(1.79±0.25 mm,P=0.038).Bland-Altman plots of both APAC eyes and PACS eyes showed small mean difference without the presence of proportional bias.However,compared with PACS eyes,the Bland-Altman plot of APAC eyes had wider 95%LoA and more outliers outside the 95%LoA.CONCLUSION:Despite the small ACD values,the two methods in measuring ACD,Lenstar and UBM,demonstrate substantial consistency when measuring PACS eyes.However,for the APAC population,the agreement might be limited,and thus the measured values should be interpreted with caution.
文摘目的比较lenstar LS 900与角膜超声测厚仪测量人中央角膜厚度(CCT)的差异,为临床检查提供参考。方法选取屈光手术者26例(52只眼),分别用lenstar LS 900和角膜超声测厚仪测量双眼CCT,两种仪器均测量3次或以上,取其平均值,对所得数据进行统计学分析。结果 52只眼CCT测量结果:lenstar LS 900为(538.27±40.38)μm,A超为(534.15±41.12)μm。两种测量方法结果差异有统计学意义(t=2.9643,P<0.005)。但测量结果高度相关(r=0.970),前者所测CCT较后者厚(4.11±10.11)μm。结论 lenstar LS 900与角膜超声测厚仪对CCT的测量结果呈较高的相关性,两种仪器具有较高的测量准确性,lenstar LS 900可以作为角膜厚度的测量工具之一。