期刊文献+
共找到1篇文章
< 1 >
每页显示 20 50 100
Forearm Loss Caused by Automated Non-Invasive Blood Pressure Cuff Malfunction: A Hearsay Report
1
作者 Steven M. Shulman Yunseok Namn +1 位作者 stanislav lando Patrick Discepola 《Open Journal of Anesthesiology》 2015年第11期227-232,共6页
Failure of an automated blood pressure cuff to deflate when a patient is under general anesthesia can lead to catastrophic consequences if unnoticed for more than three hours [1]. We present this as a hearsay case in ... Failure of an automated blood pressure cuff to deflate when a patient is under general anesthesia can lead to catastrophic consequences if unnoticed for more than three hours [1]. We present this as a hearsay case in which an automated blood pressure cuff of the Spacelabs Ultraview Clinical Workstation monitor (model No. 90385) applied pressure for about five hours resulting in limb thrombosis. In order to analyze this catastrophe, simulation scenarios were tested to elucidate the possible errors and malfunctions that may have led to this injury. We present the analysis of the advantages and validity of the hearsay case report. We also include our proposed criteria that should be required when a hearsay case is considered for publication. 展开更多
关键词 Monitor MALFUNCTION HEARSAY COMPARTMENT Syndrome NON-INVASIVE Blood Pressure
暂未订购
上一页 1 下一页 到第