Difference between low-volume and high-volume Andersen samplers in measuring atmospheric aerosols 被引量:1
1
作者 Fengfu Fu Nobuo Shinohaya +3 位作者 mitsuo ito Xueqin Xu Mincong Shen Liangjun Xu 《Particuology》 SCIE EI CAS CSCD 2008年第3期218-222,共5页
The mass concentration and size distribution of aerosols in Tokaimura were investigated using a high-volume and a low-volume Andersen sampler. A difference was found using the two samplers: the concentration of total... The mass concentration and size distribution of aerosols in Tokaimura were investigated using a high-volume and a low-volume Andersen sampler. A difference was found using the two samplers: the concentration of total aerosols determined with the high-volume sampler is smaller than that of the low-volume sampler by 70-90% throughout the year. Compared to the high-volume sampler, low-volume sampler gave lower concentration for aerosols 〉7 μm, higher concentration for aerosols of 3.3-7.0 μm and 〈 1.1 μm, though similar results for aerosols of 1.1-3.3 μm. The low-volume sampler was found to have better separation efficiency and higher accuracy. 展开更多
关键词 AEROSOLS Cascade impactor Andersen air sampler Atmospheric particles