摘要 分析了原有时间相关单光子计数光谱仪存在的不足,提出了改进方案。研制了高速数据采集系统,采用PCI总线技术、FPGA技术,开发了高速光谱数据采集卡,取代了原有的多道分析仪,数据采集速度达到20MB/s,比原有仪器提高了约200倍;改善了仪器的性能、减小了体积。介绍了光谱仪系统的集成和工作流程,并对仪器的性能进行分析,通过多种标准样品的试验数据分析和对比,光谱仪系统具有最高的灵敏度-单光子计数,测得荧光寿命可达到ps量级,而且可以测得时间分辨光谱。 The paper proposes a performance improving scheme for the conventional time-correlated single photon counting spectrometer and develops a high speed data acquisition card based on PCI bus and FPGA technologies.The card is used to replace the multi-channel analyzer to improve the capability and decrease the volume of the spectrometer.The process of operation is introduced along with the integration of the spectrometer system.Many standard samples are measured.The experimental results show that the sensitivity of the spectrometer is single photon counting,and the time resolution of fluorescence lifetime measurement can be picosecond level.The instrument could measure the time-resolved spectroscopy.
出处 《强激光与粒子束》 EI CAS CSCD 北大核心 2011年第12期3417-3420,共4页 High Power Laser and Particle Beams
基金 国家自然科学基金项目(69827002)
关键词 时间相关 单光子计数 数据采集 PCI总线 荧光寿命 time-correlated single photon counting data acquisition PCI bus fluorescence lifetime
  • 相关文献

参考文献5

二级参考文献23

  • 1张秀峰,宋克菲,于涛,韩炳冬,盛翠霞,林久令.皮秒时间相关单光子计数光谱仪的核心技术[J].光谱学与光谱分析,2004,24(12):1489-1493. 被引量:3
  • 2刘立新,屈军乐,林子扬,陈丹妮,郭宝平,牛憨笨.荧光寿命成像及其在生物医学中的应用[J].深圳大学学报(理工版),2005,22(2):133-141. 被引量:12
  • 3林久令.时间相关单光子计数光谱方法──Ⅰ.时间统计特性[J].光谱学与光谱分析,1996,16(6):14-17. 被引量:4
  • 4Becker W,Bergmann A,Koenig Tirlapur K U.Picosecond fluorescence lifetime microscopy by TCSPC imaging[C] //Proc of SPIE.2001,4262:414-419.
  • 5Becker W,Bergmann A,Biskup C,et al.Multiwave length TCSPC lifetime imaging[C] //Proc of SPIE.2002,4620:79-84.
  • 6Sch(o)nle A,Glatz M,Hell S W.Four-dimensional multiphoton microscopy with time-correlated single photon counting[J].Appl Opt,2000,39:6306-6311.
  • 7Paul P H, Lcc M P, Hanson R K. Molecular velocity imaging of supersonic flows using plused planar laser-induced fluorescence[J]]. Opt Lett, 14 : 417--419.
  • 8Reisel J R, Carter C D. Elnstein coefficients for rotational lines of the (0,0) band of the NO A^2Σ^+ ←X^2Ⅱ system[J]. J Quant Spectrose Transfer, 1992, 47(1): 43--54.
  • 9Rothe E W, Andresen P. Application of tunable excimer lasers to combustion diagnostics : a review[J]. Appl Opt, 1997, 20:3971--4033.
  • 10Andresen P, Meijer G, Schlüter H,et al. Fluorescence imaging inside an internal combustion engine using tunable excimer lasers[J]. Appl Opt, 1990, 29:2392--2404.